STIM
1. Dual channel: individually isolated circuits.
2. Amplitude: 0-90 mA; indication only; actual mA will tend to be less than indicated due to electrode impedance.
3. Type: Constant Current.
4. Waveform: Asymmetrical, rectangular bi-phasic with zero DC current.
5. Selectable pulse width: 50µS – 450µS [2% accuracy].
6. Pulse Rate selection: in the continuous mode 2 – 100 Hz [2% accuracy].
7. Time duration of the treatment selectable: 1 minute to 90 minutes.
8. Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2 volts the battery symbol will flash on/off once every second.
9. Open Electrode Detect: If an open circuit is detected at the output of channel A or B the output current will be reset at zero.
10. Ramp up time 0.3 - 9.9 seconds.
11. If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn on.
12. Physical dimensions: 108 x 62 x 23 mm.
13. Weight: 0.07KG without battery, 0.1KG with battery.
14. Environmental conditions for storage & transport:
-10 to +50 degrees Centigrade
0-90% Humidity.
All our products are manufactured in compliance with the Medical Device Directive MDD93/42/EEC under the supervision of SGS, Notified Body number 0120.